Procedimientos recomendados para el control de calidad de IMRTen tomoterapia

  1. Ruth Rodríguez Romero 1
  2. Carlos Martín Rincón 2
  3. Sara Moral Sánchez 3
  4. Patricia Sánchez Rubio 1
  5. David Sevillano Martínez 4
  1. 1 Hospital Universitario Puerta de Hierro Majadahonda, Madrid
  2. 2 Hospital Universitario de Salamanca
    info

    Hospital Universitario de Salamanca

    Salamanca, España

    ROR https://ror.org/0131vfw26

  3. 3 Hospital Universitario de La Ribera (Alzira) Valencia
  4. 4 Hospital Universitario Ramón y Cajal, Madrid
Journal:
Revista de Física Médica

ISSN: 1576-6632

Year of publication: 2018

Volume: 19

Issue: 2

Pages: 73-102

Type: Article

More publications in: Revista de Física Médica

Abstract

The tomotherapy treatment unit has exclusive and specific modalities of intensity modulated radiotherapy (IMRT) owing to its design. The increasing implantation of these kind of devices, has motivated to resume the work, initiated in the framework of the task group on IMRT of the Medical Physics Spanish Society, to provide a set of recommended procedures for the quality assurance of IMRT in tomotherapy. The tests described in this document come mainly from those proposed by the manufacturer (later compiled in the report of the American Association of Medical Physics AAPM TG-148), others are based on publications of other users, and some have been contributed by the authors themselves. In any case, for each test, one or more methodologies (including the analysis) are described in detail, indicating the required equipment (preferably the one supplied with the treatment unit), the recommended periodicity and tolerance, as well as the time and minimum staff needed.